7.26
Note:
• Selection of “idms” sets the IRMA TruPoint serial data transfer (i.e.,
direct serial connection, external modem, or connection to a network
server such as the Lantronix device) baud rate to 19200. Configure idms
for a 19200 baud rate.
• Selection of “host” sets the IRMA TruPoint serial data transfer (i.e.,
direct serial connection, external modem, or connection to a network
server such as the Lantronix device) baud rate to 9600. Configure the
host for a 9600 baud rate.
• The
Results Transfer settings are ignored when Internal Modem or LAN
10/100 are selected as the communication method.
• The Automatic Results Transfer option is not available when User
Selects is configured as the Communications Method.
Automatic Results Transfer
IRMA TruPoint analyzers that are powered by AC adapter may be configured
to automatically transfer test results to idms or another connected system.
There are two Automatic Results Transfer settings:
•
Off:
The Automatic Results Transfer feature is disabled.
•
On:
IRMA TruPoint results will automatically be transferred to idms or
other connected system when certain conditions are met (see Section 4-
Automatic Results Transfer for details). Configure Automatic Results
Transfer as follows:
1. Highlight the “on” or “off” button and press
next
to setup the Automatic Results Transfer
feature.
Automatic Device Updates (via DeviceSet)
DeviceSet is a PC software application that provides an easy and efficient way
to create, modify, restore and manage configuration settings on all IRMA
TruPoint analyzers at a site. Settings profiles are established in DeviceSet and
may be downloaded to IRMA TruPoint serially, via modem, or over a
network. A settings profile is a collection of IRMA TruPoint settings that
includes most of the IRMA TruPoint settings that may be established
manually via the IRMA TruPoint SETTING OPTIONS menu. There are 2
Automatic Device Update settings:
•
Off:
The IRMA TruPoint Automatic Device Update feature is disabled.
•
On:
IRMA TruPoint analyzers can automatically receive profile updates
from DeviceSet, and software and language updates when available.