ITC IRMA TruPoint Blood Analysis System User Manual
Page 86

5.8
SCREEN MESSAGE REASON
FOR
MESSAGE
CORRECTIVE ACTION
D
ATA TRANSFER
“Unable to Transfer - No
test results to upload”
A. send was pressed when 
there are no unsent records 
in the analyzer memory. 
B. resend was pressed when 
there are no previously sent 
records in the analyzer 
memory.  
Press the appropriate button, 
depending upon the transfer status of 
the results in the analyzer memory. 
“Unable to Transfer -
Transfer unsuccessful”
Data transfer between the 
analyzer and receiving 
system was unsuccessful.  
1. Verify that receiving system is
ready (e.g., DeviceCom is running).
2. Verify that cables and adapters are
properly connected to both the 
IRMA TruPoint and receiving 
system. Use cables and adapters 
supplied by your service provider. 
3. Verify that transfer cable is
connected to correct com port 
(specified in DeviceCom setup). 
4. If transferring serially, ensure that
the IRMA TruPoint and receiving 
system baud rates match. If the 
analyzer is configured to transfer to 
“idms”, the receiving system must 
be set to a baud rate of 19200. If the 
analyzer is configured to transfer to 
“host”, the receiving system must 
be set to a baud rate of 9600. 
5. If transferring by the LAN 10/100,
make sure that the network settings 
are correct. See Section 7-
Communications. 