5.8

SCREEN MESSAGE REASON

FOR

MESSAGE

CORRECTIVE ACTION

D

ATA TRANSFER

“Unable to Transfer - No

test results to upload”

A. send was pressed when
there are no unsent records
in the analyzer memory.

B. resend was pressed when
there are no previously sent
records in the analyzer
memory.

Press the appropriate button,
depending upon the transfer status of
the results in the analyzer memory.

“Unable to Transfer -

Transfer unsuccessful”

Data transfer between the
analyzer and receiving
system was unsuccessful.

1. Verify that receiving system is

ready (e.g., DeviceCom is running).

2. Verify that cables and adapters are

properly connected to both the
IRMA TruPoint and receiving
system. Use cables and adapters
supplied by your service provider.

3. Verify that transfer cable is

connected to correct com port
(specified in DeviceCom setup).

4. If transferring serially, ensure that

the IRMA TruPoint and receiving
system baud rates match. If the
analyzer is configured to transfer to
“idms”, the receiving system must
be set to a baud rate of 19200. If the
analyzer is configured to transfer to
“host”, the receiving system must
be set to a baud rate of 9600.

5. If transferring by the LAN 10/100,

make sure that the network settings
are correct. See Section 7-
Communication
s.