3.9
Verify/Enter Cartridge Information
11. a. If the cartridge package bar code was scanned, go to step 13.
b. If manual entry, ensure that the product
type (Type) displayed on the Verify
Information screen is correct. If the
product type displayed does not match the
type of the cartridge inserted into the
analyzer, press
back
to return to the Select
Product Type screen.
12. The Verify Information screen displays the cartridge lot number (Lot) and
Cal Code from the most recent test performed on the same product type.
a. If the lot number displayed matches the lot number of the cartridge
inserted into the analyzer, press
next
. Proceed to step 13.
b. If the lot number displayed does not match the lot number of the
cartridge inserted into the analyzer, press
edit
to choose a different
previously-entered lot, or enter a new lot for the first time.
b1. To choose a previously-entered lot,
highlight the correct cartridge lot
from the picklist. Press
next
to return
to the Verify Information screen. If the
information is correct, press
next
to
continue QC test. If the information is incorrect, press
edit
to re-
enter the lot number and Cal Code.
b2. To enter a new cartridge lot for the
first time, highlight
new
from the
picklist and press
edit
.
› Enter the cartridge lot number
from the keypad. Press
next
.
› Enter the Cal Code from the
keypad. Press
next
to return to the
Verify Information screen. If the
information is correct, press
next
to
continue liquid QC test. If the
information is incorrect, press
edit
to re-enter the lot number
and Cal Code.