3.9

Verify/Enter Cartridge Information

11. a. If the cartridge package bar code was scanned, go to step 13.

b. If manual entry, ensure that the product

type (Type) displayed on the Verify
Information
screen is correct. If the
product type displayed does not match the
type of the cartridge inserted into the
analyzer, press

back

to return to the Select

Product Type screen.

12. The Verify Information screen displays the cartridge lot number (Lot) and

Cal Code from the most recent test performed on the same product type.

a. If the lot number displayed matches the lot number of the cartridge

inserted into the analyzer, press

next

. Proceed to step 13.

b. If the lot number displayed does not match the lot number of the

cartridge inserted into the analyzer, press

edit

to choose a different

previously-entered lot, or enter a new lot for the first time.

b1. To choose a previously-entered lot,

highlight the correct cartridge lot
from the picklist. Press

next

to return

to the Verify Information screen. If the
information is correct, press

next

to

continue QC test. If the information is incorrect, press

edit

to re-

enter the lot number and Cal Code.

b2. To enter a new cartridge lot for the

first time, highlight

new

from the

picklist and press

edit

.

› Enter the cartridge lot number

from the keypad. Press

next

.

› Enter the Cal Code from the

keypad. Press

next

to return to the

Verify Information screen. If the
information is correct, press

next

to

continue liquid QC test. If the

information is incorrect, press

edit

to re-enter the lot number

and Cal Code.