ITC IRMA TruPoint Blood Analysis System User Manual
Page 42
2.12
a. Result outside of reference range: If
reference ranges were established by the
QA User, and a patient test result falls
outside of the defined limits, the result will
be flagged “H” (High) or “L” (Low).
b. Result outside of reportable range: If a
measured result is outside of the IRMA
TruPoint reportable range, that result, and
any corresponding calculated results, will
be flagged “<” (less than) or “>” (greater
than).
c. Result suppressed: If a sensor errors
during the analysis phase of a patient test,
the result for that sensor, any sensors
dependent on that sensor, and any
corresponding calculated values, will be
suppressed. Suppressed results will be dashed-out on both the screen
and printout.
Note:
• pH Normalized iCa results will be reported only for pH values within the
range of 7.2 - 7.6; if pH values fall outside of this range, pH Normalized
iCa results will be suppressed.
Complete Test Information Entry (optional);
Document Patient Notes (optional)
23. Refer to Section 2-Test Information Entry and Section 2-Patient Notes
for details.
Obtain a Printout (optional)
24. a. If the Auto-print setting is “immediate”, test results will automatically
print when the analysis is complete.
b. If the Auto-print setting is “delayed”, test results will automatically
print when the
done
button is pressed from the Results screen.
c. If the Auto-print setting is “off”, a printout may be obtained by
pressing the
button.