beautypg.com

ITC IRMA TruPoint Blood Analysis System User Manual

Page 42

background image

2.12

a. Result outside of reference range: If

reference ranges were established by the
QA User, and a patient test result falls
outside of the defined limits, the result will
be flagged “H” (High) or “L” (Low).

b. Result outside of reportable range: If a

measured result is outside of the IRMA
TruPoint reportable range, that result, and
any corresponding calculated results, will
be flagged “<” (less than) or “>” (greater
than).

c. Result suppressed: If a sensor errors

during the analysis phase of a patient test,
the result for that sensor, any sensors
dependent on that sensor, and any
corresponding calculated values, will be

suppressed. Suppressed results will be dashed-out on both the screen

and printout.

Note:

• pH Normalized iCa results will be reported only for pH values within the

range of 7.2 - 7.6; if pH values fall outside of this range, pH Normalized
iCa results will be suppressed.

Complete Test Information Entry (optional);
Document Patient Notes (optional)

23. Refer to Section 2-Test Information Entry and Section 2-Patient Notes

for details.

Obtain a Printout (optional)

24. a. If the Auto-print setting is “immediate”, test results will automatically

print when the analysis is complete.

b. If the Auto-print setting is “delayed”, test results will automatically

print when the

done

button is pressed from the Results screen.

c. If the Auto-print setting is “off”, a printout may be obtained by

pressing the

print

button.