7.13
c. Select desired analyte from the picklist. If
reference ranges will be entered for multiple
analytes, select “All”. Press
edit
.
d. Enter the lower reference limit and press
next
.
e. Enter the upper reference limit and press
next
.
f. Continue for all appropriate analytes. When complete, press
done
.
3. To set Type Specific reference ranges, select
Type Specific “on” and Master “off”.
Press
next
.
a. To activate previously defined values,
press
ok
.
b. Select Reference Range Group. Press
edit
.
c. Enter Reference Range Group Name.
Press
next
.
d. Select Reference Range Type. Press
edit
.
e. Select desired analyte from picklist. If
reference ranges will be entered for multiple
analytes, select ”All”. Press
edit
.
f. Enter the lower reference limit and press
next
.
g. Enter the upper reference limit and press
next
.
h. Continue for all appropriate analytes. When complete, press
done
.
i. Repeat the steps 3c to 3g for the reference range types.
j. When complete, press
done
.
4. For Master and Type Specific defaults, select
Master “on” and Type Specific “on”.
Press
next
.
a. To activate previously defined values,
press
ok
.
b. Please follow the above Master Default steps for 2b through 2f.
c. Please follow the above Master Type Specific steps for 3b through 3j.
d. When complete, press
done
.
5. For no Reference Ranges, select Master “off”
and Type Specific “off”. Press
next
and
done
.