B.7

REPORTABLE RANGES

The default reportable ranges for each parameter are shown in Table B-8*:

Table B-8

Measured

pH

6.000 - 8.000 pH units

pCO

2

4.0 - 200.0 mmHg (0.53 - 26.66 kPa)

pO

2

20.0 - 700.0 mmHg (2.67 - 93.33 kPa)

Hct

10.0 - 80.0 % (.100 - .800 SI)

Na+

80.0 - 200.0 (mM, mEq/L)

K+

1.00 - 20.00 (mM, mEq/L)

iCa

0.20 - 5.00 mM (0.80 - 20.04 mg/dL; 0.40 - 10.00 mEq/L)

Cl

-

30.0 - 150.0 (mM, mEq/L)

BUN/urea

BUN

3 - 150 mg/dL (1.1 - 53.5 mM)

Urea

6 - 321 mg/dL (1.1 - 53.4 mM)

Glu

20 - 500 mg/dL (1.1 - 27.8 mM)

Creatinine

0.2 - 12 mg/dL(18 - 1061 µmol/L)

Lactate

0.30 -12 mmol/L (2.7 - 108 mg/dL)

Calculated

HCO

3

-

0 - 99.9 mM

TCO

2

0 - 99.9 mM

Beb ±99.9

mM

Beecf ±99.9

mM

O

2

Sat

0 - 100%

tHb

3.4 - 27.2 g/dL (2.1 - 16.9 mM)

iCa(N)

0.20 - 5.00 mM (0.80 - 20.04 mg/dL; 0.40 - 10.00 mEq/L)