7.24
Setting up communications
1. Highlight
Communications
in the DEVICE SETTINGS picklist. Press
edit
.
2. Select a data Communications Method. If
more than one method will be used, select the
User Selects
option which allows selection of a
communications method each time data
transfer is initiated. If only one method is
selected, that method will automatically be used whenever data transfer is
initiated. Press
next
.
a. If Serial is highlighted, go to step 5.
b. If Internal Modem or External Modem is highlighted, or User Selects
is highlighted on an analyzer that has Internal or External Modem as
an option, go to step 3.
c. If LAN 10/100 is highlighted, or User Selects is highlighted on an
analyzer that has LAN 10/100 as an option, go to step 4.
3. Enter the phone number of the receiving
system. Enter a comma (,) after any digits that
may result in a phone line switching delay
(e.g., after “9” is pressed to access an outside
line). Press
next
. Go to step 5.
4. The
LAN 10/100 Settings screen displays.
a. If the information displayed is correct, press
next
and go to step 5.
b. If the information is incorrect, press
edit
.
b1. The IRMA TruPoint IP Address screen
displays. Enter the IRMA TruPoint IP
Address using the format
“aaa.bbb.ccc.ddd”, where “aaa”, “bbb”,
“ccc”, and “ddd” are each a 1 to 3 digit
value ranging from 0 to 255. Press
next
.
b2. The Gateway screen displays.
Highlight the “on” button to enable
Network Gateway support, or the “off”
button to disable Network Gateway
support. Press
next
.
If “on” was selected, enter the Gateway
IP Address and press
next
.