ITC IRMA TruPoint Blood Analysis System User Manual
Page 83

5.5
SCREEN MESSAGE
REASON FOR
MESSAGE
CORRECTIVE ACTION
P
ROCEDURAL ERRORS
“Cartridge removed
prematurely”
Cartridge was removed
during a test or was not
completely inserted.
Dispose of the cartridge and retest with
a new cartridge. Fully insert the
cartridge, and do not remove cartridge
during test.
“Time for data entry
expired”
Data was not completely
entered within the allotted
time period following
cartridge insertion.
Remove and dispose of the cartridge.
Retest with a new cartridge, making
required entries within allotted time
period.
“Sample injection time
expired”
Sample was not injected
and/or test button was not
pressed within 4 minutes of
Calibration Complete
message display.
Remove and dispose of the cartridge.
Retest with a new cartridge. Inject
sample and press test within required
time period.
“Temperature Card
removed prematurely”
Temperature Card was
removed while a
temperature test was in
progress.
Press ok to return to the QC Test
Options menu. Repeat temperature test
without removing the Temperature
Card during the test.
E
NTRY ERRORS
“Lot, Product Type
mismatch”
Cartridge Lot was entered
that cannot exist for
specified product type.
Enter correct cartridge Lot and product
type.
“Lot, Cal Code mismatch” Cartridge Lot was entered
that cannot exist for the Cal
Code entered.
Enter correct cartridge Lot and Cal
Code.
“Invalid Length- Enter X -
XX characters”
The length of an entry was
invalid (too many or too few
characters). Some features,
such as Patient ID, can be
configured to require entry
of a specific number of
characters.
Note the entry length requirements
displayed on the screen. Press ok and
enter a valid number of characters.
“Invalid Format-Enter
value in format X.XXX -
X.XXX”
The format of an entry was
invalid.
Note the entry format requirements
displayed on the screen. Press ok and
enter the value in a valid format.
“Value out of range- Enter
a value from X.XXX -
X.XXX”
The value entered was
outside of the entry range.
Note the entry range requirements
displayed on the screen. Press ok and
enter a value within the entry range.