System features and setup – ITC IRMA TruPoint Blood Analysis System User Manual
Page 17
1.9
CARTRIDGE TEMPERATURE OPERATING RANGE
The IRMA TruPoint temperature operating range is 12-30°C (54-86°F).
• Room temperature cartridges that are used in environments below the
cartridge storage temperature range (e.g., CVORs below 15°C/59°F) must
be used within 4 hours of transfer from the 15-30°C storage area to the
colder area. Cartridges that were not used within the 4 hour time limit
must be returned to the 15-30°C storage area, and go through an additional
equilibration period before they can be used. Refer to Appendix B, Table
B-7 for equilibration times.
• Refrigerated cartridges may be used until their out-of-refrigerator outdate.
INSERT CHARGED BATTERY OR AC ADAPTER
INTO IRMA TRUPOINT ANALYZER
• Insert a charged battery or AC Adapter into the IRMA TruPoint analyzer.
The insertion area is located on the left side, below the carrying handle.
SYSTEM FEATURES AND SETUP
MAIN MENU ACCESS
All IRMA TruPoint test, setup, and recall options are accessed via the main
MENU. To access the main MENU:
1. Touch the right-hand edge of the screen to turn the analyzer “on.”
a. If the User ID option has not been activated, the main MENU
automatically appears at start-up.
b. If the User ID option has been activated,
the Enter ID screen appears. This option
requires the entry of a valid User ID code
before the analyzer can be used.
Following entry of a User ID, the main
MENU displays.
The main MENU also appears following
completion of a test by pressing the
done
button
or the
Menu
button from other screens.