ITC IRMA TruPoint Blood Analysis System User Manual
Page 24
1.16
Bar Code Scan Symbologies
The bar code reader is factory programmed to read the following bar code
symbologies:
Code 11
EAN/JAN
Plessey
Code 39
Interleaved 2 of 5
Telepen
Code 128
Matrix 2 of 5
UPC
Codabar MSI
Scanner Specifications
Refer to the User’s Guide that came with the bar code reader for detailed
specifications and regulatory compliance statements.
Note:
• The Welch Allyn bar code reader uses a non-laser, red LED light source.
Connecting the Bar Code Reader
1. Ensure that the IRMA TruPoint analyzer is turned “off”.
2. Connect the bar code reader cable to the round keyboard port on the back
of the analyzer.
3. The reader will be ready for use when the analyzer is powered back “on”.
Caution:
• The IRMA TruPoint analyzer can only determine the presence or
absence of a bar code reader during analyzer power-up. Do not connect
or disconnect the reader when the analyzer is “on”. Doing so may
result in damage to the reader or analyzer, or the display of incorrect
screen prompts.
Disconnecting the Bar Code Reader
1. Ensure that the IRMA TruPoint analyzer is turned “off”.
2. Disconnect the bar code reader cable from the keyboard port on the back
of the analyzer.
Mounting the Bar Code Reader
The bar code reader may be attached to the IRMA TruPoint analyzer using
the mounting clip that came with the reader.
1. Attach the clip to the IRMA
TruPoint handle as shown.
(Figure 1-6).
Figure 1-6