beautypg.com

ITC IRMA TruPoint Blood Analysis System User Manual

Page 60

background image

3.8

Select Product Type

6. Highlight the desired product type and press

next

. The Insert Cartridge screen displays.

Note:

• If only one product type was established in TEST SETTINGS/Product

Setup, the Select Product Type screen will not display. The analyzer will
assume that the product is the same type as was established in TEST
SETTINGS/Product Setup. See Section 7-Product Setup.

Open Cartridge Package

7. Check the expiration date on the cartridge package. The analyzer will not

allow testing with expired cartridges.

8. Remove the cartridge from the package and retain the package until

cartridge information is verified.

Note:

• If an expired cartridge is inserted into IRMA

TruPoint, the Lot expired message will display,
and the test cannot proceed. Remove expired
cartridge, and initiate a new test with an in-date
cartridge. If the wrong lot was selected on

the

Verify Information screen, press

ok

to return to the Verify Information

screen and press

edit

to enter or select the correct cartridge lot.

Remove Tape and Insert Cartridge into Analyzer

9. Remove the protective tape (if

applicable) from the cartridge leads.
Do not touch the cartridge leads
after the tape has been removed. Do
NOT remove the luer cap from the
cartridge. (Figure 3-1).

Figure 3-1

10. Fully insert the cartridge into the

analyzer within 15 minutes of
opening the package. (Figure 3-2)

Note:

• Do not use a cartridge if it has been

out of the package longer than 15
minutes.

• Do not reuse a cartridge once it has

been inserted into the analyzer.

Figure 3-2