beautypg.com

Task menu > calibration – Measurement Computing eZ-Analyst rev.14.1 User Manual

Page 23

background image

eZ-Analyst

878193

Menus

3-5

An Example of Auto-Ranging

Maximum FSV set at 25.7 V

Minimum FSV set at 0.2 V

Upper Limit set at 90%

Lower Limit set at 10%

Increasing Factor set at 1.5

Decreasing Factor set at 1

In this example we have set the radio button for

Current FSV instead of Minimum or Maximum

(note 2). The starting value, in the example, is 0.20 V.

In the first iteration we see

that we have a peak of 0.19 V.

This falls outside of our band of

0.02 to 0.18 V that was

established by our upper and

lower limit percentages; i.e.,

90% of the Current FSV and

10% of the FSV.

An Example of Auto-Ranging

As a result, the Current FSV is increased by a factor of 1.5 (our Increasing Factor) and the

Current FSV becomes 0.30 V. Our limits, in volts, also changed since we are now looking

at percentages of 0.30 volts instead of the same percentages of 0.20 volts.

In the second iteration of our example, we see a 0.22 volt peak. This value is within our

established limits so the Current FSV does not change.

Note 1: If the Capture Mode is the Input Channel (Trigger Mode), the Auto Range process waits for a

trigger.

Note 2: A Start FSV of Minimum or Maximum can selected instead of Current FSV, as in our example.

Minimum FSV is the default.

Task Menu >

Calibration

When calibration is performed, a signal of known Peak level [or RMS value] is supplied to a

transducer that is connected to an active input channel. An accelerometer calibrator or piston

phone is typically used to generate the calibration signal for vibration sensors and microphones,

respectively.

Examples:

Accelerometer calibrators typically make use of linear engineering

units and, as their name implies, are used for calibrating

accelerometers.

Piston phones are most often used for calibrating microphones. Piston

phones typically make use of decibel (dB) engineering units.

eZ-Analyst includes a Calibration window for selecting the channels to be calibrated and for

entering several signal-related parameters. In addition, the calibration is actually started from

the window.

This manual is related to the following products: