beautypg.com

Measurement Computing eZ-Analyst rev.14.1 User Manual

Page 22

background image

3-4 Menus

878193

eZ-Analyst

Mode

The Mode panel consists of three radio buttons, which are used to select one of the following

range modes: AutoRange(%), AutoRange(V), or ManualRange(V).

Starting FSV

The Starting FSV (Full Scale Voltage) panel consists of three radio buttons, which are used to

set the starting FSV to Maximum, Minimum, or Current.

Channel Gauges

The channel gauges, one per channel, display the instantaneous peak value as percentage or

voltage, depending on the mode that was selected. The color of the vertical bar has the

following significance:

Yellow

The signal range is from 0 to the minimum Full-Scale Voltage set in the Auto-range Setup

window.

Green

The signal range is from the minimum to the maximum of the Full-Scale Voltage set in the

Auto-range Setup window.

Red

The signal range is over the maximum of the Full-Scale Voltage set in the Auto-range Setup

window.

%, FSVF, & Factors Display

Maximum FSV is the high-end limit for the input voltage.

Note that 25.7 V is the highest possible FSV for a ZonicBook

Medallion.

Minimum FSV is the low-end limit for the input voltage. In

the figure at the right Minimum FSV is set to 0.2 volts.

HI Accept % defines the highest acceptable percentage of peak input voltage for the

selected FSV, i.e., Current, Minimum, or Maximum. Thus, if our selected FSV was 0.2 V

and we had an upper limit of 90%; then our upper limit in volts would be 0.18 V. An

example follows as to how exceeding this value causes a range adjustment.

LO Accept % defines the lowest acceptable percentage of peak input voltage for the

selected FSV, i.e., (Current, Minimum, or Maximum). Thus, if our selected FSV was 0.2 V

and we had a lower limit of 10%; then our actual low limit in volts would be .02 V.

Incr Factor (Increasing Factor) is the factor by which the Current FSV will increase,

should the peak exceed the upper limit. In the figure we see that the Increasing factor is

1.5.

Decr Factor (Decreasing Factor) is the factor by which the Current FSV will decrease,

should the peak not reach the lower limit. Keeping the decrease factor at “1” will result in

no decrease of the Current FSV. Setting the Decrease Factor to 0.8 would cause the

Current FSV to decrease to 80% of its value if the peak fell short of the low limit.

Note: These are the same values that were entered in the Vpeak column in the Channel Setup

window. The values are immediately replaced when the Auto ranging process begins.

This manual is related to the following products: