Remote Operation
Command Descriptions
3
3-23
Example
Simulate a blood pressure of 120/80 with a heart rate of 30 BPM, a gain of 50%, and a shift of
12 mmHg.
Command
makearm
bp=adams-120/80,hr=30,
envgain=50,envshift=12
Results
The Analyzer draws the makearm graph and begins
arm simulation when the NIBP monitor starts inflating
the cuff. Test results are sent to the controller when the
cuff deflates.
C
u
ffPres
C
u
ffPeak
DeflRate
DeflTime
InflRate
InflTime
TotlTime
120/
8
0
0
142
3.1
45.1
43.2
3.1
4
8
.1
(90) A
200
160
120
40
0
20
40
60
30 BPM
G 50
S 12
8
0
fcv049.eps
Returned Data
142, 3.1, 45.1, 43.2, 3.1, 48.1; varies, depending on
parameter values
MAKEARMNEO
Use this command to simulate blood pressures from the ADAMS Neonate Family target
values. Heart rate, BP, and pressure envelope gain and shift can be modified. Parameters
for this command are described below.
Parameter Definition
Range Default
Description
HR Heart
Rate
(Beats/Minute)
30 BPM to 240
BPM in 1 BPM
Steps
80 BPM
1 BPM resolution
BP Blood
Pressure
(mmHg)
ADAMSNEO-
60/30
ADAMSNEO-
80/50
ADAMSNEO-
100/65
ADAMSNEO-
120/80
ADAMSNEO-
150/100
ADAMSNEO-
200/150
ADAMSNEO-
255/195
ADAMSNEO-
120/80