Maintenance, Avoiding damage, Cleaning – Fluke Biomedical Cufflink User Manual
Page 133: Service and calibration, Packing

Maintenance, Service, and Calibration
Maintenance
4
4-3
Maintenance
The Analyzer requires little maintenance or special care; however, it is a calibrated
measuring instrument and should be treated as such. The optional carry case is
recommended for storage. It is further recommended that the storage environment be free
from vibration.
Avoiding Damage
Do not drop the instrument or subject it to any mechanical abuse that could cause a shift
in the calibrated settings.
W
Caution
To avoid
damage to the Analyzer or adverse affects on its
performance, d
o not expose the system to temperature
extremes. Ambient temperatures should remain between 0 °C
and 50 °C.
Cleaning
Clean the exterior of the Analyzer occasionally with a cloth dampened with a mild
detergent solution.
W
Caution
To avoid damage to the Analyzer or adverse affects on its
performance, do not spray liquid directly on or immerse the
unit.
Service and Calibration
If your new Analyzer fails to operate successfully or if it needs a recommended yearly
calibration, please contact the Fluke Biomedical Service Center immediately, as indicated
under Warranty and Product Support.
W
Caution
To avoid
damage to the Analyzer or adverse affects on its
performance, a
llow only qualified technical personnel to service
the Analyzer.
Packing
If repairs or calibration are required, return the Analyzer to the factory or the nearest
service center.
1. Before returning the Analyzer for factory service, contact Fluke Biomedical Service
Center for a required Return Authorization Number.
2. Provide the following information:
3. The Analyzer serial number
4. The specific steps that reproduce your problem
5. A daytime phone number
6. Your name / company
7. A fax number (if available)