Fluke Biomedical Cufflink User Manual
Page 120

Cufflink
Operators Manual
3-46
MedTester REC # 16
CHECKLIST: ADULT*NIBP
OP CODE:TEO
DEVICE INFORMATION
PHYSICAL INSPECTION
COMMENTS:
NEXT TEST DUE DATE:
USER TIME:
ELAPSED TEST TIME: 209 SECS
GENERAL CONDITION OF EQUIPMENT........PASS
LINE CORD/STRAIN RELIEF...............PASS
TRPN (POWER ON).......................REMOTE CMD, NO DATA
INDICATORS/DISPLAYS...................PASS
CONTROLS/SWITCHES.....................PASS
TUBES/HOSES...........................PASS
FITTINGS/CONNECTORS...................PASS
IDENT.................................CUFFLINK,3.00,AUSC,PUMP
LEAKTEST..............................*
LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS
MANOMETER.............................*
STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS
STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS
POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS
MAKEARM:HR=100;BP=ADAMS-120/80........*
ADULT 120/80 AT 100 BPM...............PASS
MAKEARM:HR=80;BP=ADAMS-150/100........*
ADULT 150/100 AT 80 BPM...............PASS
MAKEARM:HR=60;BP=ADAMS-200/150........*
ADULT 200/150 AT 60 BPM...............PASS
GOTOLOCAL.............................*
GTOL..................................REMOTE CMD, NO DATA
TYPE: ADULT*NIBP
MODEL:
MANF:
SN:
LOC: DNI
CN: ADULT*NIBP
DATE: 6/05/93
TIME: 15:39:31
fcv064.eps
Figure 3-10. Sentinel Adult BP Test Results