Gotolocal, Ident, Inflate – Fluke Biomedical Cufflink User Manual
Page 92

Cufflink
Operators Manual
3-18
Example
Do a global initialization.
Command globalinit
Results
All user modifications for Pop Time, RS232 port, and autosequences
are reset to factory defaults.
Returned
Data
*
GOTOLOCAL
Use this command to switch control of the Analyzer fro a serial controller to the Analyzer
keyboard. There are no parameters for this command.
Example
Return the Analyzer control to its front panel.
Command gotolocal
Results
The Analyzer logo displays, and it returns to the local mode. The front
panel keyboard is now active.
Returned
Data
*
IDENT
Use this command to show the Analyzer name and firmware version on the controller
display. There are no parameters for this command.
Example
Identify the Analyzer and firmware version.
Command ident
Results
Display of Analyzer identification and firmware version
Returned Data
CuffLink, 3.00, Pump; varies with each Analyzer, depending on the
installed firmware and options
INFLATE
Use this command to pump up the pressure inside the Analyzer. Parameters for this
command are described below.
Parameter Definition
Range
Default
PRESSURE
Pressure (mmHg)
0 to 499.75 mmHg
200 mmHg
Example
Pump Analyzer pressure up to 275 mmHg.
Command Inflate
pressure=275
Results
The Analyzer internal pump increases pressure to 275 mmHg.
Returned
Data
*