Simulating other adams family target values, Nibp monitor testing sequence – Fluke Biomedical Cufflink User Manual
Page 44

Cufflink
Operators Manual
2-16
3. Press the
Ent
key to view the Makearm graph. If the cuff pressure is not set to zero,
repeat the above procedure.
Simulating Other ADAMS Family Target Values
Simulation of any of the ADAMS Family target values follows the same procedure as
that of the 120/80 simulation:
1. From the 120/80
Makearm
display, press the
Esc
key to return to the
ADAMS
Family Target Values
screen.
2. Choose the desired target value, as indicated in Table 2-1.
3. Press the
Ent
key. The
Makearm
display reappears with a new target value listed
below
TotlTime
.
NIBP Monitor Testing Sequence
To test the accuracy of the NIBP monitor:
1. When the cuff pressure is zero, connect the cuff adapter to the Analyzer.
2. Initiate the NIBP monitor testing sequence, referring to the monitor operators manual
as needed. The following sequence occurs:
•
The cuff inflates around the mandrel.
•
The analyzer starts the peripheral pulse simulation and draws a curve
representing the cuff pressure (oscillometric technique) as the test progresses.
•
The analyzer initiates blood pressure simulation when the inflate pressure reaches
8.0 mmHg.
•
When simulation begins, a small flashing heart becomes visible to the left of
Total Test Time (
TotlTime
, on right side of graph). This represents the heart
rate. Each color change of the heart signifies one heartbeat.
•
The analyzer terminates the simulation when the pressure upon deflate reaches
6.0 mmHg.
•
The NIBP monitor interprets and displays the measured blood pressure values
and heart rate at the completion of the test.
3. Compare the monitor values with the target values displayed on the
Makearm
display.
Note
It is normal for the monitor to have a different reading from the target
value displayed on the
Makearm
display. Readings may also be different
when the monitor is in the stat mode rather than the normal (automatic) or
manual mode. The objective is that readings be consistent and repeatable.
Accuracy for the target values cannot be specified because no standard for
blood pressure measurements exists.