Fluke Biomedical Cufflink User Manual
Page 25

Introduction and Specifications
Specifications
1
1-13
Display
Alphanumeric and Graphic LCD Display
Alphanumeric Mode............................................ 8 lines by 40 characters
Graphics Mode ................................................... 64 vertical by 240 horizontal dot matrix
Illumination.............................................................. Backlight with Viewing Angle Adjustment
Displayed Graphics................................................. Dynamic real-time NIBP cuff-pressure waveform, programmed
peripheral pulse and envelope waveforms
Control Keys
Function Keys ........................................................ F1 to F5
Enter
Escape
Cursor .................................................................... Up
Down
Left
Right
Parameter Selections
Menus ..................................................................... Pull down with on-screen help
Function keys.......................................................... Software defined, F1 to F5
Storage
Environment............................................................ Store in a dry area, temperature range of 32°F to 122°F.
Inspection or maintenance during storage.............. None required
Digital Interfaces
RS232/Serial........................................................... Baud Rates: 300, 600, 1200, 2400, 4800, 9600
Stop Bits: 1, 2
Parity: Odd, Even, Off
Handshake: Xon/Xoff, RTS/CTS, None
Parallel Port (Printer) .............................................. Centronics compatible
Pulse Sync .............................................................. 0 to 5 V dc (TTL)
Target Blood Pressure Selections
Select BP
Target Value (MAP)
(mmHg)
HR
(BPM)
Waveform Type
Adult 60/30
(40)
80/50 (62)
100/65 (75)
120/80 (90)
150/100 (115)
200/150 (165)
255/195 (215)
30
40
60
120
160
200
240
Normal Sinus
Neonate 60/30
(40)
80/50 (62)
100/65 (75)
120/80 (90)
150/100 (115)
30
40
60
120
160
200
240
Normal Sinus
Arrhythmias 120/80
(90)
80
Premature Atrial Contraction
Premature Ventricular Contraction
Atrial Fibrillation
Missed Beat
Aberrant Sinus Conduction