Fluke Biomedical Cufflink User Manual
Page 83

Remote Operation
Command Descriptions
3
3-9
Parameters for this command are described below.
Parameter Definition
Range Default Description
HR Heart
Rate
(Beats/Minute)
30 BPM to 240 
BPM in 1 BPM 
Steps 
80 BPM
1 BPM resolution
BP Blood
Pressure
(mmHg)
ADAMS-60/30
ADAMS-80/50
ADAMS-100/65
ADAMS-120/80
ADAMS-150/100
ADAMS-200/150
ADAMS-255/195
ADAMS-
120/80 
PULSEID Pulse
Identification 
(milliseconds) 
ID# / WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms
ID# 3 
(720 ms) 
Pulse width affects the rise 
time of pulse 
ENVGAIN Pressure
Envelope Gain
1 to 200% 
(10 to 2000 
RVDUs) 
100% is a 
typical value
for adult 
cuffs. 
Affects pulse strength or 
amplitude 
ENVSHIFT Pressure
Envelope Shift
-100 to +100 
mmHg 
0 mmHg
Modifies existing pressure 
envelope to simulate another 
BP 
Example: 0 +10 mmHg shift 
causes 120/80 to read 
130/90 on most monitors. 
Example
Draw a pressure envelope showing a BP of 80/50 with a heart rate of 60 BPM
Command
drawenv
hx=60,bp=adams-80/50
Results
The specified pressure envelope is drawn on the 
Analyzer display: 