Fluke Biomedical Cufflink User Manual
Page 83

Remote Operation
Command Descriptions
3
3-9
Parameters for this command are described below.
Parameter Definition
Range Default Description
HR Heart
Rate
(Beats/Minute)
30 BPM to 240
BPM in 1 BPM
Steps
80 BPM
1 BPM resolution
BP Blood
Pressure
(mmHg)
ADAMS-60/30
ADAMS-80/50
ADAMS-100/65
ADAMS-120/80
ADAMS-150/100
ADAMS-200/150
ADAMS-255/195
ADAMS-
120/80
PULSEID Pulse
Identification
(milliseconds)
ID# / WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms
ID# 3
(720 ms)
Pulse width affects the rise
time of pulse
ENVGAIN Pressure
Envelope Gain
1 to 200%
(10 to 2000
RVDUs)
100% is a
typical value
for adult
cuffs.
Affects pulse strength or
amplitude
ENVSHIFT Pressure
Envelope Shift
-100 to +100
mmHg
0 mmHg
Modifies existing pressure
envelope to simulate another
BP
Example: 0 +10 mmHg shift
causes 120/80 to read
130/90 on most monitors.
Example
Draw a pressure envelope showing a BP of 80/50 with a heart rate of 60 BPM
Command
drawenv
hx=60,bp=adams-80/50
Results
The specified pressure envelope is drawn on the
Analyzer display: