Utilities – Fluke Biomedical Cufflink User Manual
Page 62

Cufflink
Operators Manual
2-34
Note
If an auto sequence is configured to print and is run before a printer is
properly connected to the Analyzer, a Printer Not Ready message
displays.If a printout is not desired, select
Edit
and make sure the
BP Results
line says
NO
.
Utilities
The auto sequence utilities, accessible from the
Utility
submenu, lets the user edit, view,
name, and print auto sequences. To access any of the utilities described below from the
Utility
submenu (the
Auto Sequence Utilities
screen), highlight the desired utility
option and press the
Ent
key. The appropriate
screen displays.
Edit
The
Edit
utility lets the user change the settings of an auto sequence.
To modify the content of an auto sequence:
1. From the
Auto Sequence Utilities
screen, highlight
Edit
and select an auto
sequence with the function keys. The three Analyzer auto sequence editing pages are
shown below:
F1
F2
F3
F4
F5
Auto-1
Pressure Test. :
YES
Leak Test..... : YES
Pop Off Test.. : YES
Print BP Results YES
Store
No/Dec
PgDn
PgUp
Yes/Inc
fcv112.eps
F1
F2
F3
F4
F5
Auto-1 2
Blood Pressure HR (BPM) CYCLES
(1)
60/ 30 ( 40) Adult
40 1
(2) 80/ 50 ( 62) Adult
60 1
(3) 100/ 55 ( 75) Adult
80 1
(4) 120/ 80 ( 90) Adult
80 1
Store
No/Dec
PgDn
PgUp
Yes/Inc