Mkarr_pac, Mkarr_pvc, Mkarr_as – Fluke Biomedical Cufflink User Manual
Page 105

Remote Operation
Command Descriptions
3
3-31
MKARR_PAC
Use this command to set blood pressure simulation to the Premature Atrial Contraction
Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are no
parameters for this command.
Example
Simulate a premature atrial contraction.
Command mkarr_pac
Results
The Analyzer runs premature atrial contraction at 120/80 with the
heart rate set to 80 BPM
Returned
Data
*
MKARR_PVC
Use this command to set blood pressure simulation to the Premature Ventricular
Contraction Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are
no parameters for this command.
Example
Simulate premature ventricular contraction.
Command mkarr_pvc
Results
The Analyzer runs premature ventricular contraction at 120/80 with
the heart rate set to 80 BPM.
Returned
Data
*
.
MKARR_AS
Use this command to set blood pressure simulation to the Aberrant Sinus Conduction
Arrhythmia at the fixed BP and heart rate of 120/80 and 80 BPM. There are no
parameters for this command.
Example
Simulate aberrant sinus conduction.
Command mkarr_as
Results
The Analyzer runs premature aberrant sinus conduction at 120/80
with the heart rate set to 80 BPM.
Returned
Data
*