Printing documents, Printing blood pressure test results – Fluke Biomedical Cufflink User Manual
Page 65

Operation
Printing Documents
2
2-37
Init
The
Init
utility lets the user reset an auto sequence to factory default values.
To reset an auto sequence to factory default values:
1. From the
Auto Sequence Utilities
screen, highlight
Init
and select an auto
sequence with the function keys. The Analyzer warns that the auto sequence is about
to be reset.
2. Press the
Ent
key to reset the values or press the
Esc
key to return to the
Utilities
submenu without resetting.
Print All
The
Print All
utility lets the user print the content of all auto sequences.
To print the content of all auto sequences:
1. From the
Auto Sequence Utilities
screen, highlight
Print All
and press the
Ent
key. The Analyzer displays Printing all Auto Sequences while the documents are
printing.
Init All
The
Init All
utility lets the user reset all auto sequences to factory defaults.
To reset all auto sequences to default values:
1. From the
Auto Sequence Utilities
screen, highlight
Init All
and press the
Ent
key. The Analyzer warns that all auto sequences are about to be reset.
2. Press the
Ent
key to reset all values to the default or press the
Esc
key to return to
the
Auto Sequence Utilities
screen without resetting. Refer to Figure 2-11 for a
list of auto sequence defaults.
Printing Documents
Test results can be printed from the Analyzer to any parallel printer suitable for use with
a PC-compatible computer. Connect the printer cable to the port labeled
Printer
on the
Analyzer rear panel.
Note
If the user tries to print a document when the printer is the wrong type, out
of paper, not on line, or not properly connected, the Analyzer displays an
error message stating Printer not ready.
Printing Blood Pressure Test Results
To print the results of blood pressure tests:
1. Use the arrow keys to choose one of the selections under
Select BP
on the Analyzer
Main Menu
and press the
Ent
key.
2. Press
F3 Print
to enable printing. A Printing enabled message displays.
3. Choose the blood pressure value to simulate and press the
Ent
key.
4. Start the NIBP monitor to begin the test. The Analyzer sends the test results to the
printer when the cuff has completely deflated. Figure 2-7 is a sample printout.