Operation
Printing Documents
2
2-39
1. Highlight
Auto
on the
Main Menu
by pressing a
Left or Right arrow key.
2. From the
Auto
submenu, highlight
Utility
and press the
Ent
key.
3. Highlight
and press the function key below the auto sequence to be printed.
For example, to print the content of auto sequence #3, highlight
and press
F3
.
To print content of all the auto sequences, select
Print All
and press the
Ent
key.
Printing Auto Sequence Test Results
To print auto sequence test results:
1. Highlight
Auto
on the
Main Menu
by pressing a
Left or Right arrow key.
2. From the
Auto
submenu, highlight
Utility
and press the
Ent
key.
3. Highlight
Edit
and press the function key below the auto sequence to be printed. The
Auto-1
screen displays:
F1
F2
F3
F4
F5
Auto-1
Pressure Test. : YES
Leak Test..... : YES
Pop Off Test.. : YES
Print BP Results
YES
Store
No/Dec
PgDn
PgUp
Yes/Inc
fcv115.eps
4. Use the arrow keys to move the cursor to the
Print BP Results
line.
5. Press
F4
to enter a
YES
and then press
F3
to store the change.
6. If a printout of an auto sequence is not desired, press
F5
to enter a
NO
on this line
and then press
F3
to store the change.
Note
Before starting an auto sequence that will print, make sure the printer is
ready or a Printer not ready message displays, and the auto sequence does