2-6
Preliminary Procedures
The analyzer simulates a human arm and produces target value blood pressures for the
purpose of testing the accuracy of blood pressure readings on an NIBP monitor. The test
is initiated from the
Select BP
Main Menu
option that provides the parameters listed in
Table 2-1.
Table 2-1. Blood Pressure Parameters Provided by the Select BP Option
Adult Neonate
Adult
Arrhythmias
Blood
Pressure
(mmHg)
Mean Arterial
Pressure (mmHg)
Blood Pressure
(mmHg)
Mean Arterial
Pressure (mmHg)
60/30
80/50
400/65
120/80
150/100
200/150
255/195
40
62
75
90
115
165
215
60/30
80/50
400/65
120/80
150/100
40
62
75
90
115
Heart Rate Selections
Premature Atrial
Contractions
Premature
Ventricular
Contraction
Atrial fibrillation
Missed Beat
Aberrant Sinus
Conductions
Blood Pressure
and Heart Rate
30
40
60
80
120
160
200
240
Blood Pressure
fixed at 120/80 (90)
Heart Rate fixed at
80 BPM
Assembling Equipment
The following equipment is needed to test NIBP monitor blood pressure readings. Figure
2-2 shows the proper way to combine Adult cuff mandrels to simulate various sizes of
arms.
•
NIBP monitor
•
NIBP Analyzer
•
Mandrel (supplied with the Analyzer)
•
BP cuff
•
Hoses to attach the cuff to the monitor
•
Cuff adapter for the monitor DUT (device under test) (supplied with the Analyzer)