Fluke Biomedical Cufflink User Manual
Page 26

Cufflink
Operators Manual
1-14
Preprogrammed Pulse Envelope
Horizontal Axis (cuff pressure) ............................... 0 to 300 mmHg in 1.0 mmHg steps
Vertical Axis ............................................................ 2.0 mmHg (nominal)
100% gain with normal adult cuff
0 to 2% selectable
Pulse Amplitude (adult) .......................................... 2.0 mmHg @ MAP 100% gain
Repeatability ...........................................................
±
1 % of selected target value
Pulse Waveforms
Pulse ID #
Pulse Width (ms)
Rise Time (ms)
0 800
270
1 500
165
2 250 85
3 720 90
4 230 80
5 280 96
6 350
100
7 480
108
8 980
180
9 1980
460
10 1480 330
Digital Manometer
Pressure Range ...................................................... Maximum = 499.75 mmHg
Measurement Parameters ...................................... Instantaneous and peak
Pump....................................................................... 2.0 liter/minute minimum (free flow)
Automated Leak Testing
Start Pressure ......................................................... Maximum = 499.75 mmHg
Elapsed Time .......................................................... Fixed at 60 seconds
Leak Rate Range .................................................... 0.25 to 499.75 mmHg/minute
Pump....................................................................... 2.0 liter/minute minimum (free flow)
Monitor Pop-off Relief Valve Testing
Measurement Parameters ...................................... Instantaneous and peak pressure
Maximum Pressure ................................................. 499.75 mmHg
Accuracy
Dynamic NIBP Response Repeatability
(Systolic/Diastalic/Mean).........................................
±
1 % of Target Value
Cuff Pressure ..........................................................
±
1.0 % of reading (±1 mmHg)
Input Overpressure Limit.........................................
±
1500 mmHg