Remote Operation
Programming with Analyzer Commands
3
3-53
DEMO ROOM
Page: 1
Date: 06/05/93
Time: 12:34 AM
Proc. Name
SENT*ARR
Description
GENERAL CONDITION OF EQUIPMENT
LINE CORD/STRAIN RELIEF
TRPN (POWER ON)
INDICATORS/DISPLAYS
CONTROLS/SWITCHES
TUBES/HOSES
FITTINGS/CONNECTORS
IDENT
LEAKTEST
LEAKTEST: APPLY 200 MMHG (<10/MIN)
MANOMETER
STATIC PRESSURE CHECK: APPLY 200 MMHG
STATIC PRESSURE CHECK: APPLY 100 MMHG
POP OFF/RELIEF VALVE: APPLY >250 MMHG
MKARR_PAC PRE. ATRIAL CONTRACTION
AARRHYTHMIA PAC 120/80 AT 80 BPM
MKARR_PVC PRE VENTRICULAR CONTRACTION
AARRHYTHMIA PVC 120/80 AT 80 BPM
MKARR_AF ATRIAL FIBRILLATION
AARRHYTHMIA ATRIAL FIB 120/80 AT 80 BPM
GOTOLOCAL
GTOL
5
GEN/PHYS
GEN/MECH
REMOTE
GEN/ELEC
GEN/ELEC
GEN/PLMB
GEN/PHYS
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
AUTO SEQ
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
medTester remote command
medTester autosequence
Type
medTester Control
0.00
PM Labor Time
Relational Checklist Report
Listed In Order Of Relational Procedure
fcv079.eps
Figure 3-17. Sentinel Adult Arrhythmic BP Checklist