Fluke Biomedical Cufflink User Manual
Page 124

Cufflink
Operators Manual
3-50
MedTester REC # 18
CHECKLIST: NIBP*NEO
OP CODE: TEO
DEVICE INFORMATION
PHYSICAL INSPECTION
COMMENTS:
NEXT TEST DUE DATE:
USER TIME:
ELAPSED TEST TIME: 263 SECS
GENERAL CONDITION OF EQUIPMENT........PASS
LINE CORD/STRAIN RELIEF...............PASS
TRPN (POWER ON).......................REMOTE CMD, NO DATA
INDICATORS/DISPLAYS...................PASS
CONTROLS/SWITCHES.....................PASS
TUBES/HOSES...........................PASS
FITTINGS/CONNECTORS...................PASS
IDENT.................................CUFFLINK,3.00,AUSC,PUMP
LEAKTEST..............................*
LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS
MANOMETER.............................*
STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS
STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS
POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS
MAKEARMNEO:HR=60;BP=ADAMSNEO-150/100.....*
ADAMS NEONATE 150/100 AT 60 BPM.......PASS
MAKEARMNEO:HR=120;BP=ADAMSNEO-100/65..*
ADAMS NEONATE 100/65 AT 120 BPM.......PASS
MAKEARMNEO:HR=200;BP=ADAMSNEO-60/30...*
ADAMS NEONATE 60/30 AT 200 BPM........PASS
GOTOLOCAL.............................*
GTOL..................................REMOTE CMD, NO DATA
TYPE: NIBP*NEO
MODEL:
MANF:
SN:
LOC: DNI
CN: NIBP*NEO
DATE: 6/05/93
TIME: 15:44:19
fcv071.eps
Figure 3-14. Sentinel Neonate BP Test Results