Fluke Biomedical Cufflink User Manual
Page 128

Cufflink
Operators Manual
3-54
MedTester REC # 20
CHECKLIST: NIBP*ARR
OP CODE: TEO
DEVICE INFORMATION
PHYSICAL INSPECTION
COMMENTS:
NEXT TEST DUE DATE:
USER TIME:
ELAPSED TEST TIME: 208 SECS
GENERAL CONDITION OF EQUIPMENT........PASS
LINE CORD/STRAIN RELIEF...............PASS
TRPN (POWER ON).......................REMOTE CMD, NO DATA
INDICATORS/DISPLAYS...................PASS
CONTROLS/SWITCHES.....................PASS
TUBES/HOSES...........................PASS
FITTINGS/CONNECTORS...................PASS
IDENT.................................CUFFLINK,3.00,AUSC,PUMP
LEAKTEST..............................*
LEAK TEST: APPLY 200 MMHG (<10/MIN)...PASS
MANOMETER.............................*
STATIC PRESSURE CHECK: APPLY 200 MMHG.PASS
STATIC PRESSURE CHECK: APPLY 100 MMHG.PASS
POP OFF/RELIEF VALVE: APPLY >250 MMHG.PASS
MKARR_PAC PRE. ATRIAL CONTRACTION.....*
ARRHYTHMIA PAC 120/80 @ 80 BPM........PASS
MKARR_PVC PRE VENTRICULAR CONTRACTION*
ARRHYTHMIA PAC 120/80 @ 80 BPM........PASS
MKARR_AF ATRIAL FIBRILLATION..........*
ARRHYTHMIA ATRIAL FIB 120/80 @ 80 BPM.PASS
GOTOLOCAL.............................REMOTE CMD, NO DATA
GTOL..................................REMOTE CMD, NO DATA
TYPE: NIBP*ARR
MODEL:
MANF:
SN:
LOC: DNI
CN: NIBP*ARR
DATE: 6/05/93
TIME: 15:49:44
fcv077.eps
Figure 3-18. Sentinel Adult Arrhythmic BP Test Results