Testing the rs232 port and connections – Fluke Biomedical Cufflink User Manual
Page 60

Cufflink
Operators Manual
2-32
Table 2-12. RS232 Settings
Parameter Value
Baud Rate
300
600
1200
2400
4800
9600
Bits/Character
7
8
Parity Mode
Even
Odd
None
Stop Bits
1
2
Flow Control
Xon/Xoff
RTS/CTS
None
3. If saving the new values is not desired, press the
Esc
key. If saving the new values is
desired, press the
Ent
key. In either case, the Analyzer returns to the
Comm
submenu.
Testing the RS232 Port and Connections
To test the RS232 port and its connections:
1. From the
Comm
submenu, highlight
Comm Test
and press the
Ent
key. The
Comm Test
screen displays:
F1
F2
F3
F4
F5
*** Comm port test ***
Baud, Len, Parity,Stop: 9600,8,none,1
Flow Control: XON/XOFF
Transmit Receive
Hit F2 to send Hit F4 to clear
CuffLink Test this box
Transmit Clear
fcv108.eps
At the top of the screen are the current settings for the RS232 port.
2. Press
F2 Transmit
to transmit a signal to whatever is being used as the controller.
The phrase Analyzer Test should appear on the controller's monitor.