Pumppcb, Rcuserenv, Rdclock – Fluke Biomedical Cufflink User Manual
Page 110: Rdenvgain

Cufflink
Operators Manual
3-36
PUMPPCB
Use this command to determine if a pump circuit board is installed in the Analyzer. There
are no parameters for this command.
Example
Check for pump PCB.
Command pumppcb
Results
The analyzer scans for the board.
Returned
Data
Pump PCB installed or Pump PCB not installed
RCUSERENV
Use this command to copy the user envelope data from EEROM to RAM. There are no
parameters for this command.
Example
Recall user envelope data.
Command
rcuserenv
Results
Cufflink writes the user envelope data stored in EEROM to RAM.
Returned
Data
*
RDCLOCK
Use this command to display the current time, day, and date. The data returned to the
controller is configured as hour:min:sec mo/date/yr day (0 = Sunday, 6 = Saturday).
There are no parameters for this command.
Example
Display the time and date for Monday, December 5th.
Command
rdclock
Results
The Analyzer reads the clock information.
Returned Data
13:14:12 12/05/91 1; varies, depending on the current date
RDENVGAIN
Use this command to show the pressure envelope gain value (displayed in %) on the
controller's display. There are no parameters for this command.
Example
Display the envelope gain value.
Command rdenvgain
Results
reads the envelope gain value from ROM.
Returned Data
101; varies, depending on the current setting for envelope gain