Fluke Biomedical Cufflink User Manual
Page 27

Introduction and Specifications
Specifications
1
1-15
Ranges for MAKEARM Test Values
On Line Cuff Pressure ............................................ 0.0 to 500 mmHg on display
0.0 to 300 mmHg on graph
Peak Cuff Pressure................................................. 500 mmHg peak
Inflate Time ............................................................. 0.1 to 999.9 seconds
Inflate Rate.............................................................. 0.1 to 999.9 mmHg/second
Deflate Time............................................................ 0.1 to 999.9 seconds
Deflate Rate ............................................................ 0.1 to 999.9 mmHg/second
Total Measurement Time ........................................ 999.9 seconds maximum
Heart Rate............................................................... 30
40
60
80
120
160
200
240
Analog Outputs
Cuff Pressure .......................................................... 0 to 499.75 mmHg FS, ±1.0 % of reading,
±
±
.0 mmHg (cuff)
10 mV/mmHg
Pulse Pressure........................................................ 0 to 5.0 mmHg FS, ±1.0 % of reading
1.0 V/mmHg
Mandrels
Adult........................................................................ Five interlocking plastic blocks that produce four circumferences;
a maximum cuff width of 15.25 cm:
Large Adult................................................................ 39.5 cm; use all blocks
Adult.................................................................... 33 cm; use 2 curved end blocks and 2 rectangle blocks
Small Adult.......................................................... 26.6 cm; use 2 curved end blocks and 1 rectangle block
Child.................................................................... 20 cm; use 2 curved end blocks
Neonate .................................................................. One plastic truncated cylinder that accommodates three different
circumferences: 14 cm, 10 cm, and 7.6 cm; maximum cuff width of
7.6 cm.