Selecting heart rate (htrate) – Fluke Biomedical Cufflink User Manual
Page 38

Cufflink
Operators Manual
2-10
Table 2-2. Measured Test Parameters
Abbreviation
Parameter
Unit of Measurement
CuffPres
On-line Cuff Pressure
mmHg
CuffPeak
Peak Cuff Pressure
mmHg
DeflRate Deflate
Rate mmHg/second
DeflTime Deflate
Time
seconds
InflRate Inflate
Rate mmHg/second
InflTime Inflate Time
seconds
TotlTime
Total Measurement Time
seconds
The current blood pressure target value is shown just below the list of test parameters on
the
Makearm
display and indicates which BP target value is currently being simulated
by the Analyzer.
Selecting Heart Rate (HtRate)
To select the heart rate to be tested:
1. From the
Select BP
submenu, highlight
ADAMS Adult
(or
ADAMS Neonate)
and press the
Ent
key. The
ADAMS Adult Family Target Values
screen
displays:
F1
F2
F3
F4
F5
*** ADAMS Adult Family ***
Target Values
60/30 (40)
150/100 (115)
80/50 (62)
200/150 (165)
100/80 (75)
100/80 (75)
HtRate
AdjEnv
ZeroPres
fcv100.eps
2. Highlight the blood pressure target value to be simulated.
3. Press
F1 HtRate
.
Note
The term ADAMS Family refers collectively to Adult and Neonate options.