Fluke Biomedical Cufflink User Manual
Page 123

Remote Operation
Programming with Analyzer Commands
3
3-49
DEMO ROOM
Page: 1
Date: 06/05/93
Time: 12:38 AM
Proc. Name
SENT*NEO
Description
GENERAL CONDITION OF EQUIPMENT
LINE CORD/STRAIN RELIEF
TRPN (POWER ON)
INDICATORS/DISPLAYS
CONTROLS/SWITCHES
TUBES/HOSES
FITTINGS/CONNECTORS
IDENT
LEAKTEST
LEAK TEST: APPLY 200 MMHG (<10/MIN)
MANOMETER
STATIC PRESSURE CHECK: APPLY 200 MMHG
STATIC PRESSURE CHECK: APPLY 100 MMHG
POP OFF/RELIEF VALVE: APPLY >250 MMHG
MAKEARMNEO:HR=60;BP=ADAMSNEO-150/100
ADAMS NEONATE 150/100 AT 60 BPM
MAKEARMNEO:HR=120;BP=ADAMSNEO-100/65
ADAMS NEONATE 100/65 AT 120 BPM
MAKEARMNEO:HR=200;BP=ADAMSNEO-60/30
ADAMS NEONATE 60/30 AT 200 BPM
GOTOLOCAL
GTOL
5
GEN/PHYS
GEN/MECH
REMOTE
GEN/ELEC
GEN/ELEC
GEN/PLMB
GEN/PHYS
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
AUTO SEQ
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
medTester remote command
medTester autosequence
Type
medTester Control
0.00
PM Labor Time
Relational Checklist Report
Listed In Order Of Relational Procedure
fcv073.eps
Figure 3-13. Sentinel Neonate BP Checklist