Utility functions, Set clock – Fluke Biomedical Cufflink User Manual
Page 50

Cufflink
Operators Manual
2-22
4. Press
F2 Reset
to return
Pressure actual
and
Pressure peak
to zero. If
Pressure actual
is still not zero, press
F5 Zero Pressure
.
5. Press
F3 Print
to print the test results. Table 2-5 describes the measured parameters.
Table 2-7. Measured Pop Off Test Parameters
Parameter Description
Pressure Actual (mmHg)
The current pressure in the blood pressure cuff
Pressure Peak (mmHg)
The maximum pressure in the system before
release of the monitor's emergency valve
6. Press
F4 Release Pressure
to release all pressure in the system.
7. Press the
Esc
key to quit the
Manometer
mode.
Utility Functions
The analyzer enables the user to set the clock, adjust pop time, view the logo display, and
perform system tests from the
Util
submenu:
F1
F2
F3
F4
F5
Select BP Press Util
Util Comm Auto
Set Clock
Set Clock
Pop Time
Logo
System
Set calendar clock
fcv098.eps
All of these functions are system tests or adjustments and affect only the Analyzer itself.
The NIBP monitor is not affected by the
Util
options.
Set Clock
The
Set Clock
function allows changes to be made to the clock or calendar, as shown in
the upper right corner of the logo display. The time is displayed in an hour, minutes,
seconds format, while the date is displayed in the date, day, year format.
To set the Analyzer clock:
1. From the
Util
submenu, highlight
Set Clock
and press the
Ent
key. The
Set Clock
( labeled
Set Time and Date
) screen displays: