Mkarmneo40, Mkarmneo50, Mkarr_af – Fluke Biomedical Cufflink User Manual
Page 104: Mkarr_mb

Cufflink
Operators Manual
3-30
MKARMNEO40
Use this command to set blood pressure simulation to Adams Neonate 120/80, with a
heart rate of 120 BPM. There are no parameters for this command.
Example
Simulate a specific neonate BP of 120/80 at 120 BPM.
Command mkarmneo40
Results
The Analyzer runs the Adams Neonate 120/80, heart rate is set to 120
BPM.
Returned
Data
*
MKARMNEO50
Use this command to set blood pressure simulation to Adams Neonate 150/100, with a
heart rate of 120 BPM. There are no parameters for this command.
Example
Simulate a specific neonate BP of 150/100 at 120 BPM.
Command mkarmneo50
Results
The Analyzer runs the Adams Neonate 150/100, heart rate is set to
120 BPM.
Returned
Data
*
MKARR_AF
Use this command to set blood pressure simulation to the Atrial Fibrillation Arrhythmia
at the fixed BP and heart rate of 120/80 and 80 BPM.
There are no parameters for this
command.
Example
Simulate an atrial fibrillation.
Command
mlcarr_af
Results
The Analyzer runs atrial fibrillation at 120/80 with the
heart rate set to 80 BPM.
Returned
Data
*
MKARR_MB
Use this command to set blood pressure simulation to the Missed Beat Arrhythmia at the
fixed BP and heart rate of 120/80 and 80 BPM. There are no parameters for this
command.
Example
Simulate a missed beat.
Command mitarr_mb
Results
The Analyzer runs missed beat at 120/80 with the heart rate set to 80
BPM.
Returned
Data
*