Makearm – Fluke Biomedical Cufflink User Manual
Page 94

Cufflink
Operators Manual
3-20
MAKEARM
Use this command to simulate blood pressures from the ADAMS Adult Family target
values. Heart rate, BP, and pressure envelope gain and shift can be modified. Parameters
for this command are described below.
Parameter Definition
Range Default
Description
HR Heart
Rate
(Beats/Minute)
30 BPM to 240
BPM in 1 BPM
Steps
80 BPM
1 BPM resolution
BP Blood
Pressure
(mmHg)
ADAMS-60/30
ADAMS-80/50
ADAMS-100/65
ADAMS-120/80
ADAMS-
150/100
ADAMS-
200/150
ADAMS-
255/195
ADAMS-120/80
PULSEID Pulse
Identification
(milliseconds)
ID# / WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms
ID# 3
(720 ms)
Pulse width affects
the rise time of
pulse
ENVGAIN Pressure
Envelope Gain
1 to 200%
(10 to 2000
RVDUs)
100% is a typical
value for adult
cuffs.
Affects pulse
strength or
amplitude
ENVSHIFT Pressure
Envelope Shift
-100 to +100
mmHg
0 mmHg
Modifies existing
pressure envelope
to simulate
another BP
Example: 0 +10
mmHg shift
causes 120/80 to
read 130/90 on
most monitors.