Makearm – Fluke Biomedical Cufflink User Manual
Page 94

Cufflink
Operators Manual
3-20
MAKEARM
Use this command to simulate blood pressures from the ADAMS Adult Family target
values. Heart rate, BP, and pressure envelope gain and shift can be modified. Parameters
for this command are described below.
Parameter Definition
Range Default
Description
HR Heart
Rate
(Beats/Minute)
30 BPM to 240 
BPM in 1 BPM 
Steps 
80 BPM
1 BPM resolution
BP Blood
Pressure
(mmHg)
ADAMS-60/30
ADAMS-80/50
ADAMS-100/65
ADAMS-120/80
ADAMS-
150/100 
ADAMS-
200/150 
ADAMS-
255/195 
ADAMS-120/80
PULSEID Pulse
Identification 
(milliseconds) 
ID# / WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms
ID# 3 
(720 ms) 
Pulse width affects 
the rise time of 
pulse 
ENVGAIN Pressure
Envelope Gain
1 to 200% 
(10 to 2000 
RVDUs) 
100% is a typical 
value for adult 
cuffs. 
Affects pulse 
strength or 
amplitude 
ENVSHIFT Pressure
Envelope Shift
-100 to +100 
mmHg 
0 mmHg
Modifies existing 
pressure envelope 
to simulate 
another BP 
Example: 0 +10 
mmHg shift 
causes 120/80 to 
read 130/90 on 
most monitors. 