Neonate bp checklists and test results, Remote operation – Fluke Biomedical Cufflink User Manual
Page 121

Remote Operation
Programming with Analyzer Commands
3
3-47
Neonate BP Checklists and Test Results
Figure 3-11 shows a medBase Neonate BP checklist that creates results like those shown
in Figure 3-12. Figure 3-13 shows a Sentinel Neonate BP checklist that creates results
like those shown in Figure 3-14.
DATE: 06/05/1993
CHECKLIST NUMBER: 19 (start)
Ln
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
80
1
2
37
6
7
13
15
38
39
40
41
42
43
73
67
68
69
70
71
72
50
51
----->
25
\P GENERAL CONDITION OF EQUIPMENT
\P LINE CORD/STRAIN RELIEF
\M TRPN (POWER ON)
\P INDICATORS/DISPLAYS
\P CONTROLS/SWITCHES
\P TUBES/HOSES
\P FITTINGS/CONNECTORS
\M IDENT
\M LEAKTEST
\P LEAK TEST: APPLY 200 NMHG (<10/MIN)
\M MANOMETER
\P STATIC PRESSURE CHECK: APPLY 200 MMHG
\P STATIC PRESSURE CHECK: APPLY 100 MMHG
\P POP OFF/RELIEF VALVE: APPLY >250 MMHG
\M MAKEARMNEO:HR=60;BP=ADAMSNEO-150/100
\P ADAMS NEONATE 150/100 AT 60 BPM
\M MAKEARMNEO:HR=120;BP=ADAMSNEO-100/65
\P ADAMS NEONATE 100/650 AT 120 BPM
\M MAKEARMNEO:HR=200;BP=ADAMSNEO-60/30
\P ADAMS NEONATE 60/30 AT 200 BPM
\M GOTOLOCAL
\M GTOL
through line 79 - unused -
-A 4
GEN/PHYS
GEN/MECH
REMOTE
GEN/ELEC
GEN/ELEC
GEN/PLMB
GEN/PHYS
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
AUTO SEQ
Item
Check Item
Type
CHECKLIST DESCRIPTION: NIBP*NEO
CHECKLIST CONTENTS
PAGE: 4
--- end of checklist #19 --
fcv068.eps
Figure 3-11. medBase Neonate BP Checklist