Fluke Biomedical Cufflink User Manual
Page 84

Cufflink
Operators Manual
3-10
Max
BP =
8
0/50, HR = 60 BPM
1000
Min
mm
7
1.0
0.
8
0.6
0.4
0.2
0
40
8
0
120
160
200
fcv038.eps
Returned
Data
*
Example
Draw a pressure envelope with a heart rate of 30 BPM, a blood pressure of 200/150, a 50% gain, and a 
shift of 20 mmHg. 
Command
drawenv
hr---30,bp=adams200/150,envgain=50,envshift=20
Results
The specified pressure envelope is drawn on the Analyzer display:
Max
BP = 200/150, HR = 30 BPM, gain = 50
%
, shift = 20 mmHg
500
Min
mm
63
1.0
0.
8
0.6
0.4
0.2
0
120
160
200
240
2
8
0
fcv039.eps
Returned
Data
*
                    See also other documents in the category Fluke Biomedical Equipment:                                                            
                    - 7000DP Impulse Supplement (12 pages)
 - 7000DP Impulse Getting Started (36 pages)
 - 7000DP Impulse (92 pages)
 - 06-526-2200 (26 pages)
 - 07-417 (16 pages)
 - 07-443 (14 pages)
 - 07-451 (8 pages)
 - 07-487 (12 pages)
 - 07-453 (24 pages)
 - 07-555 (8 pages)
 - 07-553 (10 pages)
 - 07-605-7777 (8 pages)
 - 07-444 (46 pages)
 - 07-618 (10 pages)
 - 07-600 (6 pages)
 - 07-591 (8 pages)
 - SigmaPace 1000 (154 pages)
 - 07-653 (8 pages)
 - 07-633 (20 pages)
 - 07-649 (14 pages)
 - 07-661-7662 (12 pages)
 - 07-645 (14 pages)
 - 10100AT (80 pages)
 - 07-644 (10 pages)
 - 18-203 (4 pages)
 - 07-621 (12 pages)
 - 07-647 (12 pages)
 - 18-207 (8 pages)
 - 18-216-1000 (10 pages)
 - 18-220 (10 pages)
 - 18-228 (8 pages)
 - 18-223 (18 pages)
 - 18-229-1313 (6 pages)
 - 18-250 (8 pages)
 - 18-252 (6 pages)
 - 18-303 (6 pages)
 - 35035 (19 pages)
 - 6000-528 (18 pages)
 - 35080B (70 pages)
 - 6000-529 (16 pages)
 - 35080M (64 pages)
 - 6000-530B (16 pages)
 - 57-436 (12 pages)
 - 57-440 (12 pages)
 - 57-402 (32 pages)
 