Fluke Biomedical Cufflink User Manual
Page 109

Remote Operation
Command Descriptions
3
3-35
Parameter Definition
Range
Default
PULSEAMP Pulse
Amplitude
(RVDU)
1 to 1000 RVDUs
1000 RVDUs
PULSEID
Pulse Identification
ID# / WIDTH
0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms
Pulse #3
(720 ms)
Pulse width affects the
rise time of the pulse.
*RVDU = Relative Volume Displacement Unit, or the volume display when the pressure engine
moves one increment (one microstep).
Example
Output a pulse using the active pressure envelope and pulse width (ID).
Command pulse
Results
No parameters were specified in this example, so Cufflink determines
amplitude by reading cuff pressure and applies it to the active
pressure envelope. Cufflink also uses the current pulse width and
drives the motor to output a single pulse and then stop. Pulse is
output at PULSE output jack on the Cufflink front panel.
Returned
Data
*
Example
Output a pulse that is 231 steps in amplitude and 23 ms wide.
Command
pulse
pulseamp=231,pulseid=4
Results
Pulse #4 becomes the active pulse. If another pulse command is
issued without specifying parameters, the Analyzer uses pulseid #4.
Pulse is output at the PULSE output jack on the Analyzer front panel.
Returned
Data
*