Simulating neonatal blood pressure – Fluke Biomedical Cufflink User Manual
Page 45

Operation
Simulating Neonatal Blood Pressure
2
2-17
Simulating Neonatal Blood Pressure
Neonatal blood pressure is simulated and the monitor is tested in the same way as for
Adult blood pressure, with appropriate cuff and value differences. Target values are also
chosen from the
Select BP
submenu.
Note
Use the neonate mandrel with a neonate cuff when simulating neonatal
blood pressure.
To simulate neonatal blood pressure:
1. From the
Select BP
submenu, highlight
ADAMS Neonate
and press the
Ent
key.
The
ADAMS Neonate Family Target Values
screen displays:
F1
F2
F3
F4
F5
*** ADAMS Neonate Family ***
Target Values
60/30 (40)
150/100 (115)
80/50 (62)
200/150 (165)
100/80 (75)
100/80 (75)
HtRate
AdjEnv
ZeroPres
fcv103.eps
2. See Table 2-1 for desired neonate target blood pressure values.
3. Use the same steps as for Simulating Adult Blood Pressure:
•
Select the heart rate
•
Adjust the pressure envelope
•
Set zero pressure
•
Run the NIBP monitor testing sequence
4. Compare the monitor values with the target values displayed on the
Makearm
display.