Readpress, Reset – Fluke Biomedical Cufflink User Manual
Page 112

Cufflink
Operators Manual
3-38
Example
Read the user envelope
Command rduserenv
Results
The Analyzer takes data for the user envelope in RAM and sends it
out the RS232 port.
Returned Data
1:100. 2:100, 3:100, 4:105, 5:110 …. 298:150, 299:140, 300:130, *
READPRESS
Use this command to show a value (in 0.25 mmHg increments) on the controller's display
for the Analyzer pressure transducer. There are no parameters for this command.
Example
Read the pressure transducer value.
Command readpress
Results
The Analyzer reads the pressure transducer value stored in EEROM.
Returned Data
123.25; varies for each Analyzer
RESET
Use this command to reset the Analyzer back to factory default values. This affects the
parameters listed below. There are no parameters for this command.
Parameter Default
Value
Pulse ID number
3
Heart Rate
80 BPM
Pressure Curve ID number
A120 80
Pressure Curve Vertical Gain
100%
Pressure Curve Horizontal Shift
0 mmHg
Example
Reset CuffLink to default values
Command Reset
Results
The Analyzer resets to factory values.
Returned
Data
*