Page 53

Operation
Utility Functions
2
2-25
Logo
The
Logo
option shows the Analyzer name, current software revision, time, and date.
To view the logo:
1. From the
Util
submenu, highlight
Logo
and press the
Ent
key. The logo is
displayed:
F1
F2
F3
F4
F5
16:15:36
09/19/07
CuffLink
3.21, Pump
Non-Invasive Blood Pressure Analyzer
fcv101.eps
2. Press the
Esc
key to return the display to the
Util
submenu.
Note
No adjustments are possible in the Logo option.
System Functions
The
System
functions, accessed from the
Util
submenu, are primarily functional self
tests, available from the System Utilities screen:
F1
F2
F3
F4
F5
*** System Utilities ***
Print Test
ROM
Checksums
Key Test
Config Init
Speaker Test
User Envelope
Display Test
Pulse
Send test string to printer
fcv110.eps
Print Test
The Print Test verifies that the Analyzer can communicate with a connected printer.
To carry out the Print Test:
1. Connect the printer to the Analyzer printer port on the rear panel and power on the
printer.
2. On the
System Utilities
screen, highlight the
Print Test
option and press the
Ent
key. The Analyzer sends a test string through the printer port to the printer. If the
Analyzer is communicating correctly, the printer should print the following each time
the
Ent
key is pressed:
Printer test message