CALPUCKPOS
Use this command to calibrate puck position. This procedure positions the puck at a
known location or reference point from which subsequent puck movement is calculated.
During this procedure, the Analyzer moves the puck in a downward direction until it
extends to its full range of movement (to the chamber bottom). Then the puck moves
up and stops as soon as it is detected by an opto sensor. The Analyzer records the
number of motor steps necessary to move the puck this distance and stores the number in
EEROM. There are no parameters for this command.
Example
Calibrate the puck position.
Command calpuckpos
Results
The Analyzer calibrates the puck. Returned data is
the value that
the Analyzer stores in EEROM
Returned Data
5681; varies with each Analyzer
DEFLATE
Use this command to release all pressure inside the Analyzer. There are no parameters for
this command.
Example
Eliminate the pressure inside the Analyzer.
Command deflate
Results
The Analyzer releases pressure.
Returned
Data
*
DRAWENV
Use this command to draw an ADAMS Adult pressure envelope on the Analyzer display,
as shown in Figure 3-1. Envelope gain, shift, and pulse width can be modified.