Adult arrhythmic bp checklists and test results, Remote operation – Fluke Biomedical Cufflink User Manual
Page 125

Remote Operation
Programming with Analyzer Commands
3
3-51
Adult Arrhythmic BP Checklists and Test Results
Figure 3-15 shows a medBase Adult BP Arrhythmic checklist that creates results like
those shown in Figure 3-16. Figure 3-17 shows a Sentinel Adult Arrhythmic BP checklist
that creates results like those shown in Figure 3-18.
DATE: 06/05/1993
CHECKLIST NUMBER: 17 (start)
Ln
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
80
1
2
37
6
7
13
15
38
39
40
41
42
43
73
44
45
46
47
48
49
50
51
----->
25
\P GENERAL CONDITION OF EQUIPMENT
\P LINE CORD/STRAIN RELIEF
\M TRPN (POWER ON)
\P INDICATORS/DISPLAYS
\P CONTROLS/SWITCHES
\P TUBES/HOSES
\P FITTINGS/CONNECTORS
\M IDENT
\M LEAKTEST
\P LEAK TEST: APPLY 200 NMHG (<10/MIN)
\M MANOMETER
\P STATIC PRESSURE CHECK: APPLY 200 MMHG
\P STATIC PRESSURE CHECK: APPLY 100 MMHG
\P POP OFF/RELIEF VALVE: APPLY >250 MMHG
\M MKARR_PAC PRE. ATRIAL CONTRACTION
\P ARRHYTHMIA PAC 120/80 @ 80 BPM
\M MKARR_PVC PRE. VENTRICULAR CONTRACTION
\P ARRHYTHMIA PVC 120/80 @ 80 BPM
\M MKARR_AF ATRIAL FIBRILLATION
\P ARRHYTHMIA ATRIAL FIB 120/80 @ 80 BPM
\M GOTOLOCAL
\M GTOL
through line 79 - unused -
-A4
GEN/PHYS
GEN/MECH
REMOTE
GEN/ELEC
GEN/ELEC
GEN/PLMB
GEN/PHYS
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
AUTO SEQ
Item
Check Item
Type
CHECKLIST DESCRIPTION: NIBP*ARR
CHECKLIST CONTENTS
PAGE: 2
--- end of checklist #17 --
fcv074.eps
Figure 3-15. medBase Adult Arrhythmic BP Checklist