Fluke Biomedical Cufflink User Manual
Page 63

Operation
Using Auto Sequences
2
2-35
fcv113.eps
F1
F2
F3
F4
F5
Auto-1 3
Blood Pressure HR (BPM) CYCLES
(5)
150/100 ( 40) Adult
120
1
(6) 120/ 80 (ASC)Arrhyth
80
1
(7) 200/150 (165) Adult
160
1
(8) 120/ 80 (PVC)Arrhyth
80
1
Store
No/Dec
PgDn
PgUp
Yes/Inc
fcv114.eps
2. Use the
Up
and
Down arrow
keys to highlight a test parameter.
3. Press
F4
or
F5
to increase or decrease the value, respectively.
4. Press
F3 Store
to save the new data.
5. Press
F1 PgUp
or
F2 PgDn
to scroll up or down, respectively, between pages.
6. Select and change parameters on the new page as described in the previous steps.
7. Press the
Esc
key to quit and return to the
Auto Sequence Utilities
screen.
View
The
View
utility lets the user see the current configuration of an auto sequence. These
displays are view only; to change settings, use the Edit utility.
To view the content of an auto sequence:
1. From the
Auto Sequence Utilities
screen, highlight
View
and select an auto
sequence with the function keys.
2. Press
F1
or
F2
to scroll up or down, respectively, between pages.
3. Press the
Esc
key to quit and return to the
Auto Sequence Utilities
screen.
Name
The
Name
utility lets the user change the eight character name of an auto sequence.
To change the name of an auto sequence:
1. From the
Auto Sequence Utilities
screen, highlight
Name
and select an auto
sequence with the function keys. A box with the current auto sequence name appears.
AUTO-1
2. Press the
Left
and
Right arrow
keys to move the blinking cursor to the character to
be changed; then use the
Up
and
Down arrow
keys to scroll through the choices.
3. If saving the new name is not desired, press the
Esc
key. If saving the new name is
desired, press the
Ent
key. In either case, the Analyzer returns to the
Utility
submenu.