Fluke Biomedical Cufflink User Manual
Page 119

Remote Operation
Programming with Analyzer Commands
3
3-45
DEMO ROOM
Page: 1
Date: 06/05/93
Time: 12:32 AM
Proc. Name
SENT*ADULT
Description
GENERAL CONDITION OF EQUIPMENT
LINE CORD/STRAIN RELIEF
TRPN (POWER ON)
INDICATORS/DISPLAYS
CONTROLS/SWITCHES
TUBES/HOSES
FITTINGS/CONNECTORS
IDENT
LEAKTEST
LEAK TEST: APPLY 200 MMHG (<10/MIN)
MANOMETER
STATIC PRESSURE CHECK: APPLY 200 MMHG
STATIC PRESSURE CHECK: APPLY 100 MMHG
POP OFF/RELIEF VALVE: APPLY >250 MMHG
MAKEARM:HR=100;BP=ADAMS-120/80
ADULT 120/80 AT 100 BPM
MAKEARM:HR=80;BP=ADAMS-150/100
ADULT 150/100 AT 80 BPM
MAKEARM:HR=60;BP=ADAMS-200/150
ADULT 200/150 AT 60 BPM
GOTOLOCAL
GTOL
5
GEN/PHYS
GEN/MECH
REMOTE
GEN/ELEC
GEN/ELEC
GEN/PLMB
GEN/PHYS
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
CL
AUTO SEQ
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
Pass/Fail
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
Pass/Fail
medTester remote command
medTester remote command
medTester autosequence
Type
medTester Control
0.00
PM Labor Time
Relational Checklist Report
Listed In Order Of Relational Procedure
fcv067.eps
Figure 3-9. Sentinel Adult BP Checklist