Working mode activating procedure – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 99

- 99 -
Standard printout
Comprises four internal sections featuring different variables. For each
variable set option YES to include it in a standard printout. If option NO is
selected, then the variable is disabled, i.e. it is not printed.
Settings for header printout, weighing printout, and footer printout are
specified in point. 15.5; below description refers to setting a printout
template for report on pipette calibration.
Content of each of reports:
−
Working mode
−
User
−
Client
−
Pipette
−
Serial no.
−
No. of channels
−
Channel no.
−
Number of measurements
−
Operation with ISO 8655
−
Start date
−
End date
−
Water temperature
−
Temperature
−
Humidity
−
Pressure
−
Z coefficient
−
Measurements and
statistics
−
Statistics
−
Status
−
Empty line
−
Dashes
−
Signature
−
Non-standard printout
24.5. Working mode activating procedure
Carrying out pipettes calibration process requires installing an adapter for
pipettes calibration. The adapter does not come standard with the balance.
The adapter for pipettes calibration is a simple supplementary device
dedicated for calibration and checking of piston pipettes.