beautypg.com

RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 177

background image

- 177 -

Example1

Command:

OMI

– give available working modes

Response:

OMI

– available working modes are given

2_” Parts counting”

number + name

4_” Dosing”

12_”Checkweighing”

OK

– command completed

Example2

Command:

OMI

– give available working modes

Response:

OMI

– available working modes are given

2

(mode number)

4

12

OK

– command completed

OMS - Set working mode

Command sets current working mode of a balance.

Format: OMS_n

Accessible responses:

OMS_OK

– command completed

OMS_E

– error occurance in course of command

performance, incorrect format or lack of the

parameter

OMS_I

– command understood but not accessible at this

moment

n – parameter, decimal value specifying working mode number. See OMI

command description for detailed information.

Example

Command:

OMS_13 – set Statistics mode

Response:

OMS_OK – Statistics mode selected

OMG - give current working mode

Command gives current working mode for a given device.

Format: OMG

Accessible responses:

OMG_n_OK

– command completed, current working mode is