RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 30
- 30 -
Using
weighing processes are within set tolerance limits, and they comply with the
requirements of the quality system accepted by an organization.
Caution: the mode operates in the weighing mode only.
Available options:
•
METHOD
It is a means for designating an accepted quality standard.
Pressing a field displays a window with a list of minimal sample
weight values stored in balance’s memory, following a criterion on
determining the minimum sample.
Adding a new method is available only from the menu level
•
MODE
Lock – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. Additionally the software disables accepting the measurement
result if it is below the minimum sample limit
Warn – on selecting this option and during the weighing process
the display indicates corresponding pictograms that inform a user
whether the weighed mass is below or above the minimum sample
limit. The user can accept a measurement that is below the
minimum sample limit, but on a printout such measurement is
preceded by an asterisk (*).
Caution:
Means of setting a method of minimum sample determining is described
in point 27.9. of this user manual:
RADWAG authorized personnel is only authorized to add new values of
minimum samples.
Weighing with application of
If during a weighing process a user wants to be informed whether a
measurement exceeds the minimum sample limit for a specific weighing
threshold, then the
additional setting of the weighing mode.
Procedure:
1. Press grey coloured workspace
2. The display indicates menu: Settings, Keys, Information, Printouts and
Profiles
3. Select menu
4. Press < Minimum sample> field
5. Below window is opened