RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 30

background image

- 30 -

Using mode guarantees that the results of completed

weighing processes are within set tolerance limits, and they comply with the

requirements of the quality system accepted by an organization.

Caution: the mode operates in the weighing mode only.

Available options:

METHOD

It is a means for designating an accepted quality standard.

Pressing a field displays a window with a list of minimal sample

weight values stored in balance’s memory, following a criterion on

determining the minimum sample.

Adding a new method is available only from the menu level

MODE

Lock – on selecting this option and during the weighing process

the display indicates corresponding pictograms that inform a user

whether the weighed mass is below or above the minimum sample

limit. Additionally the software disables accepting the measurement

result if it is below the minimum sample limit

Warn – on selecting this option and during the weighing process

the display indicates corresponding pictograms that inform a user

whether the weighed mass is below or above the minimum sample

limit. The user can accept a measurement that is below the

minimum sample limit, but on a printout such measurement is

preceded by an asterisk (*).

Caution:

Means of setting a method of minimum sample determining is described

in point 27.9. of this user manual: .

RADWAG authorized personnel is only authorized to add new values of

minimum samples.

Weighing with application of mode

If during a weighing process a user wants to be informed whether a

measurement exceeds the minimum sample limit for a specific weighing

threshold, then the mode has to be enabled in the

additional setting of the weighing mode.

Procedure:

1. Press grey coloured workspace

2. The display indicates menu: Settings, Keys, Information, Printouts and

Profiles

3. Select menu - additional options on weighing process
4. Press < Minimum sample> field

5. Below window is opened